The Hunter Lab space was developed in 1948 by R. S. Hunter as a uniform color space which could be read directly
from a photoelectric colorimeter (tristimulus method). Values in this space are defined by the following formulas:
|
 |
where
X, Y, Z: Tristimulus values of the specimen (
,
,
tristimulus values can also be used.)
X0, Y0, Z0: Tristimulus values of the perfect reflecting diffuser
|
|
For the 2˚ Standard Observer and Standard Illuminant C, the above equations would become:
|
 |
|
Color difference ∆EH in the Hunter Lab color space,
which indicates the degree of color difference but not the direction,
is defined by the following equation:
|
 |
 |
|
The Munsell color system consists of a series of color charts which are intended to be used for visual comparison
with the specimen. Colors are defined in terms of the Munsell Hue (H; indicates hue),
Munsell Value (V; indicates lightness), and Munsell Chroma (C; indicates saturation) and are written as H V/C.
|
For example, for the color with H=5.0R, V=4.0, and C=14.0, the Munsell notation would be: 5.0R 4.0/14.0
|
|
|